Automatic Bacterial Analyzer for Medical Wasterwater, High Precision, High Stability and Reliability

Product Details
Customization: Available
After-sales Service: 12 Months
Warranty: 12 Months
Gold Member Since 2023

Suppliers with verified business licenses

Audited Supplier Audited Supplier

Audited by an independent third-party inspection agency

Address
Building 41, No. 18, Dongchang Road, Suzhou Industrial Park, Suzhou, Jiangsu, China
International Commercial Terms(Incoterms)
FOB, EXW, CIF, DDP
Terms of Payment
LC, T/T, PayPal, Western Union
  • Automatic Bacterial Analyzer for Medical Wasterwater, High Precision, High Stability and Reliability
  • Automatic Bacterial Analyzer for Medical Wasterwater, High Precision, High Stability and Reliability
  • Automatic Bacterial Analyzer for Medical Wasterwater, High Precision, High Stability and Reliability
  • Automatic Bacterial Analyzer for Medical Wasterwater, High Precision, High Stability and Reliability
  • Automatic Bacterial Analyzer for Medical Wasterwater, High Precision, High Stability and Reliability
  • Automatic Bacterial Analyzer for Medical Wasterwater, High Precision, High Stability and Reliability
Find Similar Products

Basic Info.

Model NO.
AQUA-900
Display
LED Screen
Appliance
Biochemistry
Portable
Non Portable
Certification
CE, ISO9001
Type
Coliform Analyzer
Detection Limit
1mpn/Ml
Analysis Range
0-100mpn/100ml, 0-10000mpn/100ml (Expandable)
Transport Package
Wooder Box
Specification
483mm ×503mm ×177mm (W×D×
Trademark
YUAN ZHENG
Origin
China
HS Code
9027908000
Production Capacity
500

Product Description

AQUA-900 E.coli Automatic Analyzer
Automatic Bacterial Analyzer for Medical Wasterwater, High Precision, High Stability and Reliability
AQUA-900 Water Quality E.coli Automatic Analyzer is a fully automatic online analyzer designed to measure coliform bacteria in water. It is developed based on  syringe pump platform, it adopts an automatic mechanical arm in conjunction with enzyme-substrate cultivation technology.  It can achieve high-precision, low detection limits, high stability, and low maintenance online monitoring equipment of coliform bacteria.
Application
•    Wastewater treatment plants,
•    Surface water, drinking water plants,
•    Medical wastewater, etc
Features
•    Real-time data collection during the cultivation process to form cultivation curves.
•    Automatic compensation for light source drift to ensure more accurate measurements.
•    Program upgrades and measurement data export can be completed using a USB drive.
•    Multiple external interfaces, including RS485,RS232,and RJ45.
•    The analyzer is equipped with an integrated UV sterilization device for regular sterilization of pure water.
•    High sampling precision is achieved with the support of syringe pump sampling technology and overflow technology.
•    Each cultivation module has independent temperature control, enabling independent high-temperature sterilization to eliminate residual interference.
•    Independent cultivation modules allow flexible switching between total coliforms, fecal coliforms, and Escherichia coli
•    The culture medium is stored in a low-temperature sealed environment and is drawn by an automatic mechanical arm, ensuring one bottle per use to prevent cross-contamination.
Technical Specification

Analysis Method: 
Enzyme Method
Analysis Range:
0-100MPN/100mL, 0-10000MPN/100mL (expandable)
Detection Limit:
1MPN/mL
Resolution:
1%
Error:
Standard <10%; Water <10%
Repeatability:
5%
Linearity Drift:
±5% F.S.
Analysis Mode:
Continuous, up to 2 groups per hour
Sampling Mode:
Manual or Remote
Marking Method:
Manual or Remote
Dilution Mode:
Manual or Remote
Alarm:
Liquid level alarm, self-check alarm, fault alarm, over-limit alarm (one relay)
Analog Output:
4-20mA analog signal
Communication:
One RS232 or RS485 port, Standard MODBUS protocol
Storage:
Approximately 20,000 sets (automatically saved on power failure); supports USB
Printing:
Optional mini industrial thermal printer
Maintenance Cycle:
Approximately 30 days, requires 45 minutes
Display Unit:
7-inch LCD, 800*480 resolution, TFT true color touch screen
Operating Environment:
5-35°C, Humidity <90% (no condensation)
Water Sample Temperature:
5-35°C
Water Sampling Unit:
For water sample circulation and retention; automatic backwashing precision filtration system (optional)
Automatic Cleaning:
Analytical cleaning, fault cleaning, power failure cleaning, emergency stop cleaning
Dimensions:
750×530×430mm (including reagent cabinet)
Weight:
45KG (including reagent cabinet)
Power Supply:
AC 220V ± 10%, 50Hz ± 1%
Power Consumption:
150W