WDet-5000 Ammonia Nitrogen Water Quality Online Analyzer
WDet-5000 Ammonia Nitrogen Water Quality Online Analyzer is developed based on our patented Permeance Measurement Technology Platform. Combined with Constant Temperature Fiber Technology and Digestion Colorimetric Integration Technology.
WDet-5000 Ammonia Nitrogen Water Quality Online Analyzer can achieve high precision, low detection limit, high stability and low maintenance of automatic ammonia nitrogen online monitoring. The product fully meets the application of pollution sources, industrial production process, municipal wastewater, etc. It has passed the certificate of CPA and CCEP.
Features
• Permeance Measurement Technology
not be interfered by chromaticity, suspended solids and bubbles
• High reliability
strong ability to resist chromaticity, calcium and magnesium ions
• High detection precision, low LLOQ (0.01mg/L), small drift for long-term operation
• Intelligent self-test and self-diagnosis function
• Modular design, highly integrated valve terminal
each valve can be maintained, disassembled and cleaned individually
• 24hrs real-time inspection
• Equipped with alarm function, quality control function and reverse control function
Specification
Measuring method: |
Salicylic Acid Spectrophotometry |
Range: |
0.01~2/20/100mg/L (extensible) |
Indication error: |
±8%(20%F.S.) /±5%(50%F.S.) /±3%(80%F.S.) |
Repeatability: |
≤2% |
Zero drift: |
≤0.02mg/L |
Span drift: |
<1% |
Memory effect: |
20%F.S. ±0.3mg/L,80%F.S. ±0.2mg/L |
LLOQ: |
0.01 mg/L |
Reagent: |
standard reagent can make 400 samples |
Maintenance: |
<2 hours/month |
Ambient temperature: |
5ºC~45ºC |
Communication interface: |
RS485/RS232/USB interface (optional) 1x4-20mA input/output (can be expanded to 2) |
Power consumption: |
200W(220±10%VAC 50Hz), without sampling pump |
Note: |
test method refers to the latest standards of Environmental Certification |